Publication: The transmission X-ray microscope at BESSY II
Loading...
Date
2003
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
The new transmission X-ray microscope (TXM) at BESSY II is the first one using soft X-rays in the water window region at an undulator beamline. For this, the condenser monochromator has been adapted to the highly collimated X-ray beam of the undulator U41. This is done by using a rotating condenser (RK) which can match the numerical aperture of any micro objective zone plate [1]. The actual setup of the TXM is given in this report. Several beam diagnostic instruments have been incorporated into the beamline and the microscope to adjust the optical elements and to study the behaviour of the complete system. The results ofthe characterization done so far will be presented as well as first images of applications will be shown.